Glucose Assay in Serum

From Bridges Lab Protocols
Revision as of 00:55, 13 February 2026 by Tiangebu (talk | contribs) (Reading Plate Instructions)
(diff) ← Older revision | Latest revision (diff) | Newer revision → (diff)
Jump to navigationJump to search

Materials

  • FujiFilm Autokit Glucose (Cat 997-03001). Dissolve the Color Reagent in Buffer Solution (150 mL) to make a bottle of 150 mL Color Solution. The solution is stable for 2 weeks at 4C.
  • Glucose Standard Solution (200 mg/dL)
  • Serum samples (thaw on ice if from -80C)

Procedure

  • Plan out 96-well plate, including samples and space for 5 standards and a blank, in duplicates.
  • Add 100 uL of Color Solution to each well.
  • Add 0, 1, 2, 3, 4, 5 uL to each of the standard wells.
  • Add 3 uL of sample to each well.
  • Incubate at room temperature for 30 mins (or 5 mins at 37C).
  • Read plate at 505 nM and 600 nM.

Calculations

  • Use the standard curve to calculate glucose levels.

Reading Plate Instructions

  • Turn on the plate reader (switch at the back) and the computer.
  • Open ‘Soft Max Pro’.
  • Click ‘Settings’.
  • Set ‘Wavelengths’ to “2”.
  • Set ‘Lm1’ to “490”, ‘Lm2’ to “650”.
  • Put plate in, REMOVE THE LID.
  • Click ‘Read’.
  • Reading: top numbers are 490 uL, bottle numbers are 650 uL.
  • To export files: click ‘file’ - ‘export’ - ‘save to computer’ - ‘Bridges’ folder - place in USB.